bokomslag ICRP Publication 97
Kropp & själ

ICRP Publication 97

Icrp

Pocket

2239:-

Funktionen begränsas av dina webbläsarinställningar (t.ex. privat läge).

Uppskattad leveranstid 10-15 arbetsdagar

Fri frakt för medlemmar vid köp för minst 249:-

  • 58 sidor
  • 2006
High-dose-rate (HDR) brachytherapy is a rapidly growing technique that has been replacing low-dose-rate (LDR) procedures over the last few years in both industrialised and developing countries. It is estimated that about 500,000 procedures (administrations of treatment) are performed by HDR units annually. LDR equipment has been discontinued by many manufacturers over the last few years, leaving HDR brachytherapy as the major alternative. 



HDR brachytherapy techniques deliver a very high dose, of the order of 1.6–5.0 Gy/min, so mistakes can lead to under- or over dosage with the potential for clinical adverse effects. More than 500 HDR accidents (including one death) have been reported along the entire chain of procedures from source packing to delivery of dose. Human error has been the prime cause of radiation events. In the present report, the International Commission on Radiological 

Protection concludes that many accidents could have been prevented if staff had had functional monitoring equipment and paid attention to the results. 



Since iridium has a relatively short half-life, the HDR sources need to be replaced approximately every 4 months. Over 10,000 HDR sources are transported annually, with the resultant potential for accidents; therefore, appropriate procedures and regulations must be observed. 



A number of specific recommendations on procedures and equipment are given in this report. The need for an emergency plan and for practising emergency procedures is stressed. The possibility of loss or theft of sources must be kept in mind.
  • Författare: Icrp
  • Format: Pocket/Paperback
  • ISBN: 9780080446585
  • Språk: Engelska
  • Antal sidor: 58
  • Utgivningsdatum: 2006-03-01
  • Förlag: SAGE Publications Ltd